Calculator features for carbohydrates, blood glucose (BG) corrections, and Insulin on Board; large 992 mm display screen for less scrolling; reminders for meal bolus and blood glucose (BG) check; waterproof tested to 12 feet/24 hours; ezBolus™, Audio Bolus, and Combo Bolus; step-by-step priming for safety; precise basal rate adjustment in 0.025 U increments to enhance fine-tuning; interchangeable covers in various colors; 24-hour, toll-free customer service staffed by healthcare professionals; PC download with ezManager® Plus software
Animas® IR 1200 Product Specifications
Size:
2.9" x 2.0" 0.76"
Weight:
3.13 oz. with battery and full cartridge
Warranty:
4 years
Motor:
DC motor with gear-reducer
Power Source:
One AA lithium (1.5V), 6–8 week life; can use AA alkaline (shorter life)
Pump Type:
Self-priming, piston-driven, disposable cartridge
Supports U-100 insulin
Cartridge Capacity:
200 units of U-100 (2.0 mL)
Waterproof:
Tested at 12 feet for 24 hours (IPX8)
Infusion Set:
All commercially available sets with standard luer lock connection
Basal Rate Profile:
Up to 12 basal rates in 4 personalized programs, plus a temporary basal rate; the duration of temporary rates can be adjusted from 30 minutes to 24 hours in 30-minute intervals and 10% increments
Basal Range:
0.025–25 U/hr
Smallest Bolus:
0.05 U
Memory:
600 Boluses, 120 TDD, 30 Alarm, 60 Prime, 30 Suspend, 270 Basal Records
Occlusion Alarm:
Patented sensor and algorithm to detect occlusions promptly
Over-Delivery Alarm:
Over-infusion prevented by continuous self-tests, layer upon layer of redundancies, and numerous other safeguard systems
Near Empty Alarm:
User set: 10, 20, 30, 40, or 50 U